马尔文仪器采用了很多先进的技术,比如激光散射,Zeta电位测量,非侵入型后散光学和图像分析
Support

您的位置主页SupportValidation

Validation Overview

What is Validation?

The short textbook definition of validation is: "Proof of Specified Performance". It follows that a validation process requires; Specification, Performance and Means of Proof.

Who is responsible for validation?

The company using the process or system to be validated. There is no such thing as a validated system which can be delivered into a user company without any additional validation work being required. The reason for this is that the most important part of the validation process is one of determining and proving the suitability of a purchased system for its designated function and, going forward, its continued appropriate use by the purchaser. The ease of being able to carry out this process will define the validatability of the system. This will depend on the quality of the supplier's supporting documentation and procedures.

How is it done?

The first step in the validation process is the production of a validation plan where a company sets out the intended validation procedures for a given process or system. This plan should describe the validation steps to be taken and define successful outcomes at each stage of the process. The most common structure of the validation process is based on the performance of the following five Qualification steps:

  Specification Qualification.
 
  Construction Qualification.
 
  Installation Qualification.
 
  Operational Qualification.
 
  Performance Qualification.
 
 

How does Malvern Instruments help users to validate their systems?

By having a system of Lifecycle documentation for hardware and software development which conforms very closely to GAMP guidelines.By offering a level of OQ documentation which sets the standard for analytical equipment.

By offering QSpec validation contracts to provide a ready framework for system validation.

By the provision of Quality Audit Standards and latices for ongoing verification of the performance of systems and sample dispersion units.

What about 21 CFR part 11 compliance?

Malvern's program of remediation of its major systems to meet the requirements of 21 CFR Part 11 is now well under way with compliance-enhancing versions of software now being available for the Mastersizer 2000 and the Spraytec. These versions of software provide multi-layer access, audit trails, Windows NT® -type security features such as password aging and are widely recognized as providing a very high standard of compliance.

发电子邮件给同事,向他们介绍物质表征和流变技术的资料库